Veterinary Biopsy Forceps
Dual-function forceps: biopsy and foreign body removal for precise tissue sampling
Product Basic Information
| Product Name | Veterinary Biopsy Forceps |
| Standard | Q/SCZL 007-2026 |
| Registered Entity | Shanghai Shenchuang Zhilian Trading Co., Ltd. |
| Device Type | Sterile Disposable Veterinary Interventional Surgical Instruments |
| Sterilization Method | Ethylene Oxide Sterilization |
Full Specifications
| Jaw Diameter | 0.5mm~3.0mm |
| Total Instrument Length | 500mm~3000mm |
| Product Categories | Sampling forceps / Foreign body forceps |
| Pliers jaw hardness | 290Hv~310Hv |
| Riveted/Welded Tensile Strength | ≥80N |
| Clamp jaw surface roughness Ra | ≤0.8μm |
| Main Material | Medical-grade stainless steel |
| Sterilization Method | Ethylene Oxide Sterilization |
| Packaging Specifications | 1 pcs/box |
Compatible Animals
Applicable Scenarios
Categorized into biopsy forceps and foreign body forceps, these instruments are designed for endoscopic procedures in the digestive, respiratory, and urinary tracts of livestock (dogs, cats, pigs, cattle, sheep) and companion animals. Biopsy forceps collect tissue samples from lesions or mucosal areas for laboratory pathology analysis. Foreign body forceps are used to grasp and remove ingested foreign objects, small stones, or detached tissues from body cavities.
Not applicable:Not suitable for grasping hard large bones, sharp rigid metal foreign bodies, or deep large masses.
Product Performance Features
- 1Overall surface must be clean, free of oil stains, black spots, cracks, and sharp edges. It should be smooth with no rust or burrs to prevent scratching the animal's mucosal lining.
- 2The plier head opens and closes smoothly without sticking. All riveted and welded joints are secure, withstanding a tensile force of 80N without weld failure or fracture.
- 3Jaw hardness: 290Hv~310Hv; moderate strength allows complete tissue sampling without deformation.
- 4Low surface roughness, corrosion-resistant, passes neutral salt spray testing, and resists rusting to prevent wound contamination during surgery.
- 5Sterilized and sterile at the factory; meets hygiene requirements for minimally invasive endoscopic surgery.
- 6Compact design adapts to various cavity lengths and supports a wide range of veterinary endoscopy procedures.
Standard Operating Procedure
- 1Pre-operative inspection: Verify that the outer packaging is intact, unbroken, airtight, and dry; check that the instrument has not expired. After opening the package, inspect the clamping body for deformation, cracks, or corrosion.
- 2Device Insertion: Coordinate with veterinary endoscopy to gently advance the biopsy forceps catheter into the animal's lumen until it reaches the target tissue or foreign body. Do not force insertion.
- 3Sampling (Forceps): Push the handle to open the jaws, position against the lesion tissue, pull back the handle to clamp, and gently extract a small sample of the pathological tissue.
- 4Object Retrieval (Foreign Body Forceps): Open the jaws to envelop the object, then close them securely to prevent slippage.
- 5Extraction: Withdraw the instrument slowly and steadily to retrieve tissue samples or foreign bodies. Handle gently throughout to minimize mucosal injury.
- 6Postoperative Handling: This device is for single use only. Dispose of it according to veterinary medical waste regulations after use. Do not clean, re-sterilize, or reuse.
Safety Precautions
- Intended solely for biopsy sampling and foreign body removal in the body cavities of livestock and companion animals. Strictly prohibited for use in non-veterinary clinical settings.
- Do not use if the packaging is damaged, damp, expired, or if the tool body is deformed or rusted.
- Do not bend or pull the tube forcefully, as this may cause the jaw tip to fracture at the weld and leave metal fragments inside the animal.
- If the jaw jams, fails to open/close, slips during grasping, or if the instrument body cracks during surgery, stop immediately and replace with a new instrument.
- Do not reuse. Repeated use can compromise the mechanical properties of the pliers and lead to cross-contamination.
- Do not force oversized or excessively hard foreign objects into the jaws to prevent permanent deformation or breakage.
Storage and Transportation Requirements
Storage Conditions
Store in a clean, well-ventilated, dry, and light-protected area. Keep away from corrosive, toxic, flammable, or explosive materials. Do not store outdoors or expose to sun and rain.
Shipping Requirements
Handle with care. Protect against rain, moisture, sun exposure, and crushing. Do not transport with toxic, hazardous, or sharp items to prevent puncturing the sterile packaging.
Expiration date
2